Equivalence test: Scaled parameter range

Add one or several equivalence tests for the scaled range of the Test or Control sample parameter to the Standard sample parameter (divided by the scaling parameter) to your assay. To add the tests as assay suitability tests, add them as children of the Assay suitability tests element. To add the tests as sample suitability tests, add them as children of the Sample suitability tests element.

Important: Test definitions are ignored without notification if the parameter you select results in invalid test definitions (for example, if a test cannot be performed for the model you use).

The following table lists all available tests and indicates which models and assay elements they support:

Equivalence test on the scaled range of the: Linear parallel-line model 4-parameter logistic model 5-parameter logistic model 3-parameter logistic model with fixed lower or upper asymptote Slope ratio model Assay elements
Upper asymptote (A) (nonlinear models) no yes yes with fixed Lower asymptote no Test, Control
B parameter (nonlinear models) no yes yes yes no Test, Control
Lower asymptote (D) (nonlinear models) no yes yes with fixed Upper asymptote no Test, Control
Asymmetry parameter (G) (nonlinear models) no no yes no no Test, Control
Intercept (linear models) no no no no yes Test, Control
Slope (linear models) yes no no no yes Test, Control
Difference of asymptotes (A - D) (nonlinear models, Standard only) no yes yes no no Standard, Test, Control

This element has the following attributes:

Data type Multiplicity Usage Default value
Node 0...unbounded optional none

Context:

/Quantitative response assay/Analysis/Suitability tests/Assay suitability tests/alternatives/

/Quantitative response assay/Analysis/Suitability tests/Sample suitability tests/alternatives/

/Test system definition (Quantitative response assay)/Suitability tests/Assay suitability tests/alternatives/

/Test system definition (Quantitative response assay)/Suitability tests/Sample suitability tests/alternatives/