Gain optimization

The gain of a plate reader reflects its dynamic range, indicating the level at which reasonable measurements can still be carried out. bioMérieux endotoxin detection assay documents let you define a gain optimization routine, therefore providing the data you need to adjust the signal amplification to maximize dynamic range and ensure the results remain within the instrument's detection limits.

Note: You can run gain optimization independent from other analytical settings, that is, gain optimization does not require a full standard curve or a fitted model.

The dynamic range of a plate reader is the ratio between the brightest and darkest light intensity. By optimizing the gain, you can avoid over-saturating the detector with signals that are too strong or failing to detect signals that are too weak. This ensures that both high and low concentration samples are measured accurately, contributing to reliable endotoxin quantification​.

Figure 1. Gain optimization routine of a bioMérieux endotoxin detection assay document

Instrument gain

Use the Creatable elements pane to add an Instrument element, and define the name and gain of your instrument.

Gain optimization

Use the Creatable elements pane to add a Gain optimization element, and define the optimization parameters.

Parameter Description

Used standard step

Select the dose step of the Standard sample to be used as reference for determining the optimum instrument gain. Commonly, the dose step of the Standard sample with a nominal concentration of about 0.5 EU/ml is used for determining the optimum instrument gain.

Current gain

Select the gain used for the measurement. This is the gain you entered for the instrument in the Documentation section of this assay document.

Optimum dRFU value

Enter the value expected to be measured for the concentration of the selected dose step. The value to use depends on the analytical configuration and the test kit and instrument used for the assay.

Scaling factor

Enter a value for the scaling factor. The scaling factor is used to determine the optimum instrument gain, taking into account the Optimum dRFU value and the average dRFU of the dose step. The value to use depends on the analytical configuration and the test kit and instrument used for the assay.

Tip: You can add a test on the maximal percentual CV of the Standard sample responses to your assay. If the CV of the selected step exceeds the maximum CV, the result of the gain optimization is invalidated.